Click-On Inline Glass Indicating Moisture Trap
Moisture in carrier gas lines will prematurely degrade oxygen and hydrocarbon traps
and increase detector noise. As a precaution, we higly recommend installing a moisture
trap before the hydrocarbon and oxygen traps on all carrier gas lines.
Click-On Inline Glass Indicating Oxygen Trap
Oxygen is a column killer. It is present even in UHP gases, as minutes leaks at
fittings allow oxygen to influx against the concentration gradient. Because oxygen
can enter a gas line at any fitting or during gas bottle exchange, the oxygen trap
should be the last connection before the gas line enters the chromatograph.
Click-On Inline Glass Hydrocarbon Trap
Use a hydrocarbon trap if your gas has a potential source of hydrocarbon contaminants
or if you suspect you are observing carrier gas ghost peaks. Install the hydrocarbon
trap after the moisture trap, to prevent moisture from degrading the hydrocarbon-trapping
ability of the activated carbon in the hydrocarbon trap.
Click-On Inline Glass Indicating Hydrocarbon Trap
The hydrocarbon trap is also available with AT indicator.
Click-On Inline Glass Indicating Combi (Moisture + Oxygen) Trap
The Combi trap is ideal for purifying carrier gas. It contains oxygen and moisture
scrubbers in one easy to change economical cartridge
Click-On Inline Glass Indicating Triple (Moisture + Hydrocarbon
+ Oxygen) He Specific Trap
The Helium-Specific Triple trap is ideal for purifying helium in GC/MS systems.
This trap is packed and purged under helium and contains oxygen, moisture and hydrocarbon
scrubbers in one cartridge.
Click-On Inline Glass Indicating Triple (Moisture + Hydrocarbon
+ Oxygen) He Specific Trap
Install an indicator after the Click-On inline filter (using the Click-On double
Connector) so there is no confusion about when to replace traps
Click-On Inline Glass Indicating Combi (Moisture + Hydrocarbon)
Trap
The Fuel Gas Inline trap is perfect for purifying flame ionization detector (FID)
fuel gases, removing both moisture and hydrocarbons. Using the Fuel Gas Filter for
FID Hydrogen and air will produce a stable baseline, improving overall reproducability
and sensitivity. Ideal for use in combination with a zero air generator.
Click-On Inline Glass Indicating Triple (Moisture + Hydrocarbon
+ Oxygen) H2 Specific Trap
The Hydrogen-Specific Triple trap is ideal for purifying hydrogen in GC/ MS systems.
This trap is packed and purged under hydrogen and contains oxygen, moisture and
hydrocarbon scrubbers in one cartridge.
Click-On Inline Moisture Trap
Moisture in carrier gas lines will prematurely degrade oxygen and hydrocarbon traps
and increase detector noise. As a precaution, we higly recommend installing a moisture
trap before the hydrocarbon and oxygen traps on all carrier gas lines.
Click-On Inline Oxygen Trap
Oxygen is column killer. It is present even in UHP gases, as minutes leaks at fittings
allow oxygen to influx against the concentration gradient. Because oxygen can enter
a gas line at any fitting or during gas bottle exchange, the oxygen trap should
be the last connection before the gas line enters the chromatograph.
Click-On Inline Hydrocarbon Trap
Use a hydrocarbon trap if your gas has a potential source of hydrocarbon contaminants
or if you suspect you are observing carrier gas ghost peaks. Install the hydrocarbon
trap after the moisture trap, to prevent moisture from degrading the hydrocarbon-trapping
ability of the activated carbon in the hydrocarbon trap.
Click-On Inline Combi (Moisture + Oxygen) Trap
The Combi trap is ideal for purifying carrier gas. It contains oxygen and moisture
scrubbers in one easy to change economical cartridge.
Click-On Inline Triple (Moisture + Hydrocarbon + Oxygen) Trap
The Triple trap is ideal for purifying carrier gas. It contains oxygen, moisture
and hydrocarbon scrubbers in one easy to change economical cartridge.
Click-On Inline Triple (Moisture + Hydrocarbon + Oxygen) He Specific
Trap
The Helium-Specific Triple trap is ideal for purifying helium in GC/MS systems.
This trap is packed and purged under helium and contains oxygen, moisture and hydrocarbon
scrubbers in one cartridge.
Click-On Inline Combi (Moisture + Hydrocarbon) Trap
The Fuel Gas Inline trap is perfect for purifying flame ionization detector (FID)
fuel gases, removing both moisture and hydrocarbons. Using the Fuel Gas Filter for
FID Hydrogen and air will produce a stable baseline, improving overall reproducability
and sensitivity. Ideal for use in combination with a zero air generator.
Click-On Inline Triple (Moisture + Hydrocarbon + Oxygen) H2 Specific
Trap
The Hydrogen-Specific Triple trap is ideal for purifying hydrogen in GC/ MS systems.
This trap is packed and purged under hydrogen and contains oxygen, moisture and
hydrocarbon scrubbers in one cartridge.
Click-On Connector Stainless Steel
Includes a set of two Stainless Steel Click-On Connectors.
Click-On Connector Brass
Includes a set of two Brass Click-On Connectors.
Click-On Double Connector Stainless Steel
Click-On Double Connector to connect Stainless Steel trap with Glass Indicating
trap
Wall mount bracket set
Wall mount bracket clamps for Click-On Cartridges.
Click-On O-ring replacement set
A set of 10 + 10 O-rings to replace the O-rings on your Click-On connectors.